EUROPE


Vibration Controller

K2 Sprint

K2 Sprint

Common hardware supports all types of vibration test.
While inheriting all of the performance and features of K2, K2 Sprint offers improved cost-effectiveness with 2-channel hardware.
K2 Sprint is best-suited to single monitor-channel operation.
[Differences from K2]
●Input 2 channels (No expansion)
●Output 2 channels (No expansion)

Features

Features

■User-friendly interface

Instinctive Operation

[Intuitive operation]

Easily-recognised icons are used for file management.

Simplified definition

[Short-cut operations]

The key operations are initiated using a simplified set of commands.

Help

[Scheduler]

Define the test schedule and execute it in order.

■Easy-to-use functions

Web monitor

[Web monitor]

A user can monitor a vibration test remotely by connecting over the internet.

Report generator

[Report generator]

Results can be exported to Microsoft Word in a range of customised formats.

■Precision control capabilities
K2 employs high-resolution 24-bit A/D and D/A converters to achieve high-precision, repeatable control.
■Safety interlock function
A hardware mute function operating independently of software provides complete protection for the vibration system and payload.
■E-mail notification
Remote sites can be notified by email of any abnormality during testing or termination of a test.
■DataViewer
Software to display result data file saved after the test of SINE, RANDOM and SHOCK.

Software

Software

Common optional software Outline
CAPTURE:
Analogue waveform
signal data program

Provides analogue waveform signal capture, saved data can then be used as the reference of
SHOCK, BMAC waveform controls or Random vibration PSD control.

• Sampling Frequency 51.2kHz maximum
• Data Length Maximum 5000k points
• Input Channel Maximum 64
• Waveform edit/analysis function Filtering, Frequency transfer processing, PSD transfer,
Transmissibility ratio between channels
SCHEDULER:Test scheduler Pre-defined tests can be executed in sequence.

Specifications

Main Enclosure
AC Power Single-phase AC, 100V-240V (Auto-selected)
External Communication Contact I/O (for Emergency stop)
Ambient Conditions 0-40℃, below 85% RH, no condensing
Dimensions W345 x H40 x D210mm (excluding connectors etc)
Mass Approximately 6.0kg
●Input Section
Number of Channels 2
Input Connector BNC
Input Signal
Charge, Voltage, IEPE
Charge Amplifier Sensitivity 1.0mV/pC or 10mV/pC
Charge Amplifier Cut-off 0.32Hz
Maximum Input Charge Input ±10000pC or ±1000pC
Voltage Input ±10000mV
IEPE Input ±10000mv
Sampling Frequency 51.2kHz maximum
Voltage Input Coupling AC or DC
AC Coupling Cut-off 0.1Hz
CCLD Amplifier(IEPE) +24VDC,3.5mA
TEDS(IEPE) Version0.9,Version1.0
A/D Converter Type : ΔΣ
Resolution : 24-bit
Dynamic Range: 117dB
Digital Filter : Pass-band ripple ±0.001dB
: Stop-band attenuation 110dB
●Output Section
Number of Channels
2 (a channel is occupied for drive output)
Output Terminal BNC
Output Signal
Voltage
Maximum Output
±10000mV
Sampling Frequency
51.2kHz maximum
D/A Converter Type : ΔΣ
Resolution : 24-bit
Dynamic Range: 120dB
Digital Filter : Pass-band ripple ±0.005dB
: Stop-band attenuation 75dB

Minimum Specification of PC

● One built-in PCI Express x1 slots
● Microsoft Windows 10 Pro ( 64bit )、Windows 10 IoT Enterprise ( 64bit ) or
 Windows 7 Professional SP1 ( 32bit/64bit ) is required*.
● Memory required ( for 8 input channels )
 Windows 7 ( 32 bit ) :2 GB or more
 Windows 7/10 ( 64 bit ) :4 GB or more
● DVD-ROM Drive ( required for installation )
● One USB port
● Resolution of monitor and PC required 1280 x 1024 or more
* Recommended OS and memory vary depending on software, options , number of I/O channels, etc.